Type 1250 Sodium Flame Photometer
Specification and Services.
Purpose
To measure the concentration of a standard Na.Cl. aerosol both before and after it has passed through an air filtration device.
Sensitivity.
Better than 5ng/m³ giving a minimum detectable penetration of less than 0.0005%. With a challenge aerosol of 13mg/m³.
Maximum Concentration.
Continuous indication of mass concentrations up to 13mg/m³, but for short periods up to 20mg/m³ if calibrated to this level.
Sample Flow Rate
2 l / min +/- 0.1 l / min
Gas Supply.
Hydrogen 600 ml / min @ 2 bar. [when compressed gas cylinders are used a suitable 2-stage regulator is required with the maximum out put pressure of 5 bar.]
Air Supply.
Minimum flow 50 l/min @ 5 bar.
Electrical Power.
110 to 264 Vac. 40 to 400 Hz. Single Phase.
Dimensions.
0.58m long x 0.5m high x 0.29m deep.
Weight.
24 Kgs.
Finish.
Case : – GRP colour Mid Grey.
Front Panel :- Semi-Gloss White.